检查和调整电子水平仪零位的方法 Method for checking and adjusting the zero position of an electronic level
将被检测到的电子水平仪放在已经调整到大致水平位置的平板(或机床导轨)上,靠近定位块。气泡稳定后,气泡末端的读数为a1,然后按水平方向调整180。°,准确放在原来的位置,根据次阅读的一侧,记下气泡另一端的阅读量为a2,两次阅读量差的一半为零位偏差=(a1-a2)/2格。如果零位偏差已经超过许可范围,则需要高速零位机构。另一种方法是找到相对零位点,而不是调整零位。
Place the detected electronic level on a flat plate (or machine guide rail) that has been adjusted to a roughly horizontal position, close to the positioning block. After the bubble stabilizes, the reading at the end of the bubble is a1, and then adjust 180 horizontally. Accurately place it in its original position, and based on the side of the second reading, record the reading volume of the other end of the bubble as a2. Half of the difference between the two readings is the zero position deviation=(a1-a2)/2 grid. If the zero position deviation has exceeded the allowable range, a high-speed zero position mechanism is required. Another method is to find the relative zero point, rather than adjusting the zero point.
Place the detected electronic level on a flat plate (or machine guide rail) that has been adjusted to a roughly horizontal position, close to the positioning block. After the bubble stabilizes, the reading at the end of the bubble is a1, and then adjust 180 horizontally. Accurately place it in its original position, and based on the side of the second reading, record the reading volume of the other end of the bubble as a2. Half of the difference between the two readings is the zero position deviation=(a1-a2)/2 grid. If the zero position deviation has exceeded the allowable range, a high-speed zero position mechanism is required. Another method is to find the relative zero point, rather than adjusting the zero point.